Ultra series pure(Tap water inlet)
Ultra series pure water system (Tap water inlet)
Specification
|
Standard
|
Eliminating endotoxin
|
Low TOC
|
Comprehensive
|
Model
|
Ultra pure
|
Ultra pure UF
|
Ultra pure UV
|
Ultra pure UVF
|
Ultra pure plus
|
Ultra pure plus UF
|
Ultra pure plus UV
|
Ultra pure plus UVF
|
Flow procedure
|
PF+AC+RO+DI+TF
|
PF+AC+RO+DI+UF+TF
|
PF+AC+RO+UV+DI+TF
|
PF+AC+RO+UV+DI+UF+TF
|
Application
|
Microanalysis Environmental analysis
|
Molecular biology
|
Micro organic analysis
|
Molecular biology
|
AA,ICP,IC
|
PCR, gene research
|
Environmental analysis
|
Micro organic analysis
|
Buffer disposing
|
Pharmacy research
|
HPLC,TOC
|
Environmental analysis
|
Pharmacy research
|
Medicine examining,
|
VOC,GC/MS
|
Pharmacy research
|
Medicine examining,
|
Cell cultivating
|
Pharmacy research
|
Medicine examining
|
|
IVF etc.
|
Medicine examining,
|
Cell cultivating, IVF etc.
|
Feed water requirement
|
Tap water: TDS<200ppm,5-45°C,1.0-4.0Kgf/cm2
|
(if inlet TDS>200ppm,extra pretreatment is recommended)
|
Purification system
|
Pretreatment unit
|
Special spun fiber filter ("plus" model: outside 10" spun fiber filter)×1+Special active carbon block filter×1+Special active carbon block filter×1
|
RO unit
|
100GPD RO membrane×1 ("plus" model:2×100GPD RO membrane)
|
Subsequent unit
|
Ultrapure polishing resin cartridge× 4+(0.45+0.2)µm terminal filer×1
|
UV model:+Double wavelength(185&254)nm UV cartridge×1
|
UF model:+5000 Doulton UF cartridge×1
|
UVF model:+Double wavelength(185&254)nm UV cartridge×1+5000 Doulton UF cartridge×1
|
Ultrapure water quality
|
Resistivity
|
18.2MΩ.cm@25°C
|
Heavy metal
|
<0.1ppb
|
TOC
|
<10 ppb
|
<5 ppb
|
Endotoxin
|
-
|
<0.001EU/ml
|
-
|
<0.001EU/ml
|
particle(>0.22µm)
|
<1/ml
|
Bacteria
|
< 1cfu/ml
|
RO water quality
|
< TDS(ppm,mg/l) of tap water×5%
|
Output(25°C)
|
15 liters/hour* ("plus" model:30 liters/hour*)
|
Flow rate
|
2 Liters/min(with pressure tank)(Less output with UF/UV cartridge)
|
Pure water outlet
|
RO water ,ultrapure water
|
Control system
|
Automatic electronic pressure sensor controlling, RO membrane auto flushing, automatic stop without water, automatic stop when water tank full, automatic cutting off water when pump stopping, guaranteeing 24 hours' work.
|
Water quality monitor
|
TDS test pen+ on-line resistivity monitor
|
Electric requirements
|
AC110-220V,50/60Hz, 48W/plus model:72W
|
Dimension and weight
|
Length×Width×Height:41×22×42cm / 12-14 kg
|
Standard configuration
|
Main body(including:1 set cartridge)+4.0 gallon water tank+TDS pen+accessory box
|
§ Inlet water :TDS=200ppm, 25°C, 50psi and 15% recovery rate.
§ GPD=gallon per day, 1gallon=3.78L.
§ The quality of inlet water will effect output′s and cartridge′s life.
§ PF:Pretreating, AC:Active carbon, RO:Reverse osmosis, DI:Ion exchange, UV:Ultraviolet(Double wavelength:254&185nm) UF:Ultrafiltration, TF:Terminal filter
|