Welcome to www.boynindustrial.com
   Home About us Products News Feedback Contact us  
Posted 2014-4-23 Micropipette

Product Search

 
Products
  View:Kjeldahl Nitrogen Determination System
  Graphite Digester  |    Auto distiller  |  

K1100

 K1100 Auto Kjeldahl analyzer is an automatic device integrating distillation and titration functions designed based on classic Kjeldahl nitrogen determination method. It’s equipped with the latest core control system, powerful automatic degree and high-quality components, can easily achieve automatic waste discharge and cleaning of boiling tubes and titration cups, control steam supply and have real-time detection of condensation temperature. High-accuracy charging pump and titration system ensure test results accuracy, and multiple fluid-level detection gives smooth test process.

Titration precision as high as 1.0μL/step

K1100 Auto Kjeldahl analyzer has exclusive design i.e. linear motor micro-control titration system which ensures accurate results; the external use is easy for the observation of the whole titration process; the built-in high-sensibility color sensing unit ensures accurate final point determination; extremely high-precision plunger titration unit achieves unprecedented precision as high as 1.0μL/step.

Brand new touch interaction system

The powerful calculation function based on ARM microprocessor unit can be operated directly through the touch screen; powerful detection function has all test steps under real-time monitoring with status displayed in figures; solution barrel level detection function and high-grade pump effectively ensure correct solution supply; samples can be detected and classified by the system, and traditional test methods and parameters are built in and can be freely used in the test process.

Correct steam control

New material PTC constant-temperature heating modules are used in the heating unit of the steam generation system, and the use of metallic heater speeds up the fluid boiling, ensuring the generation of a great deal of steam in a short period; digital electronic temperature measurement and mechanical temperature measurement provides the heating unit with double protections; customized steam pressure regulating valve guarantees the steam generation system safety and ensures uniform steam output; and steam flow is also adjustable with software system.

 

 

Characteristics:

1.    Automatic completion of distillation, titration, calculation, printing, waste discharge and cleaning;

2.    Automatic waste discharge and cleaning ensure operator safety and save work hours;

3.    External titration cup design gives operator real-time control of the whole test process;

4.    Multiple protections i.e. failure self-detection, hints of condensed water shortage, displacement of safety door and nitrogen tube, and so on;

5.    Steam flow is controllable, satisfying different test requirements;

6.    Regent barrel enjoys fluid absence warning function, ensuring smooth test going;

7.    High-precision charging pump and titration ensure test results accuracy;

8.    Large LCD touch screen gives visual operation and abundant information, enabling user to have a good command of it;

9.    USB or RS485 interface is optional for PC connection.

Technical data:

1.        Sample capacity:             Solid≤5.00g/sample, liquid≤20mL/sample;

2.        Measuring range:             0.1mg N - 200mg N;

3.        Analysis time:                5 - 10 min/sample;

4.        Recovery:                   ≥ 99.5;

5.        Burette volume:              1.0ul/step;

6.        Reproducibility:              Average value relative error 0.5%;

7.        Storage capacity:            1000 pieces;

8.        Interface:                   USB or RS485;

9.        Power supply:               220V 50Hz;

10.    Power:                     2KW;

11.    Water consumption in the distillation process: 1.5L/min;

12.    Cooling water temperature:   20℃;

13.    Ambient temperature:        10℃ - 28℃;

Scope of application

K1100 Auto Analyzer is widely used in food processing, feed production, tobacco, livestock, soil fertility, environmental monitoring, medicine, agriculture, scientific research, teaching, quality control and other fields for the test of nitrogen or protein content, can also be used for the test of ammonium, volatile fatty acid / alkali, and so on. 

 

 

<< Back to previous page....

 
Home | About us | Products | News | Feedback | Contact us

Copyright©2010 Boyn Industrial Co., Ltd. All Right Reserved. Power by Shkj.net

ADD:  Rm.1101,Building 1,No.1428 Huanding Rd,Jianggan Dist.,Hangzhou 310021,China
Tel:+86-571-87963916 Fax:+86-571-85383683 Mobil:+86-15306522608
Skype:boyn-eric E-MAIL: boyn@boyninstrument.com ,ericboyn@yahoo.com

Skype Me